CMA 31 Linear Microdialysis Probe
Ideal for peripheral tissues such as skin, muscle, heart, adipose tissue, liver, eye, pancreas and other organs as well as tumors.
- CMA8010631 CMA 31 Linear Microdialysis Probe, 10mm membrane length, pkg. of 4
Cutaneous Implantation, Anesthetized Animal
For CMA 31 Linear Microdialysis Probe: To avoid ultra filtration over the membrane the probe should be used with a Perfusion Fluid consisting of a Ringer solution containing 30g Dextran 60/1000mL.
- Anesthetize the animal. Be sure to keep the animal’s body temperature normal during the surgical procedure and the study period by using a homeothermic blanket.
- Clean and shave the animal.
- Mark the skin with a felt tip marker where the introducer needle should go in and exit, leaving at least an 18 mm window if drug transport is studied. Tunnel the introducing needle at the appropriate depth required for the experiment.
- Load the CMA pump with a Perfusion Fluid-filled syringe (either a Luer lock single use syringe or a glass syringe with a fixed needle). Make sure there are no air bubbles.
- Connect the inlet Luer lock connector to the single use syringe or cut off the Luer lock connector from the probe inlet with a sharp cut, and connect it to the glass syringe needle via a blue Tubing Adapter. To facilitate the handling of Tubing Adaptors , they s hould be pre-s oaked in ethanol for a minimum of 10 minutes.
- Start the pump at 5 µL/ min. Confirm that liquid is flowing from the outlet of the probe. Maintain this flushing rate for approximately 2-3 min to remove air bubbles.
- Reduce the flow rate to operational rates, usually 1-3 µL/ min.
- Remove the protective tubing from the probe. Slide the probe outlet tubing through the needle end of the introducer carefully so the membrane sits in the middle of the introducer. Make sure not to kink the tubing.
- Check that the flow rate is the same as before the insertion. Hold the probe in place against the skin while carefully withdrawing the introducer needle.
- Secure the probe position with surgical tape or surgical glue.
- Wait at least 30 min for stabilization before starting the sample collection.